BIRD SA-2000A Site Analyzer
This state of the art instrument incorporates a host of advanced features and capabilities that range from producing reliable measurements to increasing technician productivity with it’s unprecedented ease of use and powerful analytical and report generating functions. The SA2000 rejects on channel interfering signals to +13dBm, so you can test antennas located in active sites. Work when it is convenient – not just when a site can be powered down.
Model SA-2000 Specifications:
Frequency Characteristics:
- Frequency Accuracy: ±150 kHz (75 ppm)
- Frequency Resolution: 50 kHz
- Frequency Range: 806-2300 MHz.
- Return Loss: 0 to –60dB Max.
- Input: +22dBm.
Features:
- Color display.
- Pass fail indication.
- Store up to 60 measurements with time/date stamp.
- Built in parallel port.
- Serial interface supports data transfer to a PC with compatible software.
- Number of Points: 238
Other Informations :
Measurement Range and Resolution:
Range: Resolution:
- Return Loss: 0.0 to -60.0 dB 0.1 dB
- VSWR: 1.00 to 99.99 0.01
Measurement Uncertainty (after 1 display refresh):
- Return Loss: Determined graphically from composite of 42dB
directivity, 1.2 source VSWR. and instrumentation error.
- VSWR: Calculated from Return Loss.
- Test Port: Connectors: N, Female normally supplied.
- Impedance: 50 Ohms.
- Directivity: 42 dB, after calibration.
- Measurement Better than 1 sweep per 4 seconds, test
Speed: port open, y-axis set to full scale.
- Immunity to _> 13 dB interferer at desired
Interfering measurement frequency.
- Signals: Maximum Input _> 22 dBm (Damage Level)
Signal: Output Power: < 0 dBm
Distance to Fault (DTF) Measurement
Range:
- X-Axis: 238 xR
- Y-Axis: 0 to -60 dB
Accuracy:
- X-Axis: ± 2% of full-scale range with Vp = 1
- Y-Axis: Same as for Return Loss Measurements
- Vp Range: 0.20 to 1.00, or 1.00 to 99 percent
Power Measure Mode:
Function: Displays power from Bird power sensors, VSWR alarm and BPM (specifications determined by sensor)
General :
Data Storage: 300 traces in fundamental data format stored
in non volatile memory recall and display traces in
any of the display formats.
PC/Remote Power Sensor Interface port:
- Connector: Female DB-9, compatible with PC serial port
- Protocol: Serial RS-232,9600 baud, 8 data bits, 1
stop bit, no parity, and no handshake.
Maintenance:
- Printer Interface Port:
- Connector: Female DB-25. compatible with PC parallel port.
- Compatibility: HP Deskjet printers with PCL Level 3 protocol.
Power Requirements:
- Internal: Lithium-ion rechargeable battery 3 hours minimum
operating time
- External DC: 9 to 16 Vdc
- External AC: 90 to 264 Vac @ 45 to 66 Hz
Physical Specifications:
- Dimensions: 10.44" x 8.38" x 3.28" (265 x 212 x 83 mm)
- Weight: < 5 lbs. (2.3 Kg)
Environmental Specifications:
- Operating Temp: -10oto50°C(14etol22°F)
- Storage Temp: -40° to 80°C (-40° to 176'F)
- Humidity: 95% maximum (non-condensing)
- Altitude: up to 15.000 feet (4.572 m)
International Standards:
- EMC: Conforms to: EN 63126-1:1997
- Safety: Conforms to: EN 61010-1:1993, with Amendment A2:1995